Optical Power Meter 67
B Remote Control Commands
This section presents detailed information about the commands and
queries supplied with your PM-1100 Optical Power Meter.
IEEE 488.2 Commands—Quick Reference
The Optical Power Meter recognizes the main commands identified in
IEEE-488.2. The table below summarizes these commands. The
commands are fully explained on the following pages.
Command Function
*CLS Clear status command
*ESE Standard event status enable command
*ESE? Standard event status enable query
*ESR? Standard event status register query
*IDN? Identification query
*LOK
a
a. This command can only be used with RS-232 communication.
Set Remote Lockout programming state
*OPC Operation complete command
*OPC? Operation complete query
*REM
a
Set Remote programming state
*RST Reset command
*SRE Service request enable command
*SRE? Service request enable query
*STB? Read status byte query
*TRG (Not supported)
*TST? Self-test query
*WAI (Not supported)