Remote Control Commands
Optical Power Meter 89
Product-Specific Commands—Description
SENSitivity:CORRection:OFFSet[:MAGNitude]
Description This command sets an offset value that is applied when dB or
dBm units are selected.
Syntax SENS:CORR:OFFS[:MAGN]<space><numeric_value>
[<space>DB]
Parameters The <numeric_value> is an offset with dB units in the ±99.999
format.
Example SENS:CORR:OFFS:MAGN 22.105
SENSitivity:FREQuency
Description This command selects a data sampling rate.
Syntax SENS:FREQ<space><numeric_value>
Parameters The <numeric_value> is the sampling rate with sec
-1
units.
0.1, 0.5, 1, 5, 10, 20, 40 samples per second.
Example SENS:FREQ 20
SENSitivity:FREQuency?
Description This query returns the current sampling rate.
Syntax SENS:FREQ?
Response Returns the current sampling rate in samples/second units.
Example SENS:FREQ?