Auto Tests
Copper, VDSL2, ADSL2+ and IP Triple-Play Test Set 29
Configuring CQ Auto Tests
Common Parameters
The Common Param. tab allows you to
set up parameters common to all tests
in the auto test function.
Each parameter is described below:
³ Cable Type defines the gauge of
the cable in use. If wire gauge is
measured in AWG units, the
available choices are: AUTO,
19 AWG, 22 AWG, 24 AWG, or
26 AWG. For mm gauge wire, the available choices are: AUTO,
0.32 mm, 0.40 mm, 0.50 mm, 0.60 mm, 0.80 mm, 0.90 mm, or
1.20 mm.
³ Temperature specifies the temperature of the cable under test in
either F (Fahrenheit) or C (Celsius) units.
³ Isolation Test Time(sec) specifies the duration of the isolation test.
³ Input Impedance displays the input impedance in the loop. The
available choices are: >1 MOhm, 100 KOhm.
To set parameter values:
1. Press the up/down arrow keys to highlight the desired parameter.
2. Press to display the list or select the value.
3. Press the up/down arrow keys to highlight the desired value.
OR
4. Press the left arrow key to erase the existing value, and then use the
alphanumeric keypad to enter a value. To cancel the entry, press .
5. Press to accept the value.