Vue de la page 41
Auto Tests
34 AXS-200/635i
Configuring CQ Auto Tests
Thresholds 1
The Thresholds 1 tab allows you to set
threshold values for the auto tests.
Each parameter is described below:
³ Max. AC Voltage T-R (V) defines
the maximum AC threshold
voltage for the T-R leg of the circuit.
Specify a value between 1 and 30.
³ Max. AC Voltage to GND (V)
defines the maximum AC threshold voltage to ground for the circuit.
Specify a value between 1 and 30.
³ Max. DC Voltage T-R (V) defines the maximum DC threshold voltage
for the T-R leg of the circuit. Specify a value between 1 and 400.
³ Max. DC Voltage to GND (V) defines the maximum DC threshold
voltage to ground for the circuit. Specify a value between 1 and 400.
³ Min. Resistance T-R(MOhm): defines the minimum threshold
resistance for the T-R leg of the circuit. Specify a value between
0 and 999.
³ Min. Resistance T-G(MOhm) defines the minimum threshold
resistance for the T-G leg of the circuit. Specify a value between
0 and 999.
³ Min. Resistance R-G(MOhm) defines the minimum threshold
resistance for the R-G leg of the circuit. Specify a value between
0 and 999.