Exfo RTU 310-310G IP Services Test Head Manuel d'utilisateur Page 454

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 547
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 453
Maintenance
442 RTU-310/310G
Recalibrating the Unit
Recalibrating the Unit
Manufacturing and service center calibrations are based on the
ISO/IEC 17025 Standard, which states that calibration documents must not
contain a recommended calibration interval, unless this has been
previously agreed upon with the customer.
Validity of specifications depends on operating conditions. For example,
the calibration validity period can be longer or shorter depending on the
intensity of use, environmental conditions and unit maintenance. You
should determine the adequate calibration interval for your unit according
to your accuracy requirements.
Under normal use, EXFO recommends calibrating your unit every two
years.
Vue de la page 453
1 2 ... 449 450 451 452 453 454 455 456 457 458 459 ... 546 547

Commentaires sur ces manuels

Pas de commentaire